Advanced Settings—Kits step in workflow bar
IMPORTANT! We recommend using the default settings. Consult your local Field Service Engineer before modifying advanced settings.
- In the Kits step in the workflow bar, expand the Advanced Settings box, then select Customize.
-
Edit one of the following settings.
Setting
Description
Script that the Ion Chef™ Instrument follows to perform the templating reaction. We recommend that you do NOT change this setting.
Select your forward library key, if you have one.
Enter your test fragment key.
Select one of the following base calibration options:
-
Default Calibration—allows a random subset of wells to be used for base calibration. This option uses TMAP to align the training subset of wells and is recommended if a good reference for the template is available.
-
Enable Calibration Standard—allows wells belonging to the Calibration Standard to be selected as the training subset.
-
Blind Calibration—uses the same random subset of wells as Default Calibration but does not require an alignment step to generate the calibration model. This option is recommended if the template does not align well to a reference genome or if no reference is specified.
See your template kit user guide for more details.
Select your forward 3' adapter.
Select the flow order. For more information, see flow order.
-